ClaudeCodeMod

All shelves / MCP servers

Thorlabs Optical Power Meter

io.github.K-Dense-AI/labmcp-thorlabs-pm · 19 stars · Python · Apache-2.0

MCP server MCP server for Thorlabs optical power meters (PM100 family, PM400, PM101, PM5020) over SCPI.

Install

The repo has no one-line install. Follow its README.

Published as

  • PyPIlabmcp-thorlabs-pmstdio

From the server's entry in the official MCP Registry.

Open the repo

Configuration

NameSet asWhat it is
LABMCP_ADDRESSOptionalEnv varInstrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
LABMCP_SIMULATEOptionalEnv varSet to 1 to use the built-in simulator (no hardware)
LABMCP_READ_ONLYOptionalEnv varSet to 1 to disable all state-changing tools
LABMCP_LIMITSOptionalEnv varSafety limit overrides, e.g. max_temperature_c=80

Files

servers/physics/thorlabs-power-meter/README.md

Thorlabs Optical Power Meter — MCP Server

Let an AI agent read laser power, log stability and warm-up drift, set the wavelength correction, averaging and range, and dark-zero the sensor on Thorlabs optical power meters through their documented SCPI command set over USB (USBTMC / VISA).

Package labmcp-thorlabs-pm
Instruments PM100D, PM100A, PM100USB, PM400 consoles with C-series photodiode and thermal sensors. PM101 and PM5020 use the same documented commands but are unverified (see Notes)
Interfaces USB (USBTMC, via VISA)
Protocol SCPI: PM100D Operation Manual 17654-D02 §6.4, PM100USB Operation Manual 19570-D02 §6.3, PM400 Operating Manual v1.1 §5.5, and Thorlabs' per-model SCPI command references
Status 🧪 simulated: tested against a wire-level SCPI simulator, not yet verified on hardware. Report a hardware test

Try it without hardware

uvx labmcp-thorlabs-pm --simulate --check
uvx labmcp-thorlabs-pm --simulate --option sim_sensor=S302C --check   # thermal sensor head

The simulator is a PM100D with an S120C photodiode (or S302C thermopile) looking at a 1.234 mW, 632.8 nm HeNe laser. Setting the wrong wavelength on the photodiode gives a proportionally wrong reading, as on the real bench.

Connect your power meter

  1. Driver. Thorlabs meters enumerate as USB Test & Measurement Class (USBTMC) devices, vendor ID 0x1313.
    • Windows: the Thorlabs driver (TLPM / Thorlabs VISA) is not visible to PyVISA. Install the free NI-VISA runtime, then switch the meter to the NI-VISA driver with the Thorlabs Driver Switcher (installed with the Thorlabs Optical Power Monitor software), and start the server with ?backend=@ivi (see below). This is the route Thorlabs documents for PyVISA.
    • Linux: pyvisa-py talks USBTMC directly (installed with this package). Give your user access to the device, e.g. /etc/udev/rules.d/99-thorlabs-pm.rules:
      SUBSYSTEM=="usb", ATTR{idVendor}=="1313", MODE="0666"
      
      then sudo udevadm control --reload-rules and re-plug the meter. NI-VISA also works (?backend=@ivi).
    • macOS: pyvisa-py + libusb (brew install libusb); NI-VISA is an alternative.
  2. Find the resource string: uvx labmcp ports (or python -m pyvisa info). It looks like USB0::0x1313::0x8078::P0012345::INSTR (the product ID depends on the model: PM100D 0x8078, PM100USB 0x8072).
  3. Test the connection:
    uvx labmcp-thorlabs-pm --address USB0::0x1313::0x8078::P0012345::INSTR --check
    uvx labmcp-thorlabs-pm --address "visa://USB0::0x1313::0x8078::P0012345::INSTR?backend=@ivi" --check   # NI-VISA
    For the second input of a PM5020 add --option channel=2.

Add to your MCP client

Claude Code

claude mcp add power-meter -- uvx labmcp-thorlabs-pm --address USB0::0x1313::0x8078::P0012345::INSTR

Claude Desktop / Cursor / Windsurf (claude_desktop_config.json, .cursor/mcp.json, …)

{
  "mcpServers": {
    "power-meter": {
      "command": "uvx",
      "args": ["labmcp-thorlabs-pm", "--address", "USB0::0x1313::0x8078::P0012345::INSTR"]
    }
  }
}

Add --read-only to allow readings but block wavelength, averaging, range and zero changes. For other clients, generate the snippet with uvx labmcp config thorlabs-pm --address <resource> --client vscode (also cursor, codex, claude-code).

Tools

Tool Kind Description
get_command_log 👁 read Return the most recent raw commands sent to / replies received from the instrument (newest last). Useful for debugging and for recording what was done.
get_connection_info 👁 read Report which instrument is connected (identity, address, simulated or real), whether the server is read-only, and the active safety limits. Call this first.
get_device_info 👁 read Identify the meter console and the attached sensor head (model, serial, type, wavelength range, power ranges), and report the present wavelength, averaging, range and zero value.
log_power_series 👁 read Log a series of power readings to characterise laser stability, warm-up or drift. Returns mean, stdev, min/max, RMS and peak-to-peak stability (%) and drift (%/min), plus the readings (downsampled to max_points; use save_path, a new .csv file, to keep all of them).
read_power 👁 read Read the optical power once (in W, plus dBm and a formatted string). The reading uses the configured correction wavelength and averaging; set the wavelength first.
read_sensor_temperature 👁 read Read the sensor head temperature in °C (thermal sensors and other heads with a built-in temperature sensor only).
reconnect 🛑 safety Close and re-open the connection to the instrument (e.g. after it was power cycled or a cable was re-plugged).
set_averaging 🎛 control Set how many samples the meter averages for each reading. More averaging lowers noise but slows each reading (PM100 series: ~3 ms per sample; PM400/PM101/PM5020: 1 ms).
set_range 🎛 control Select auto-ranging, or a fixed power range that fits range_w (the meter picks the most sensitive range that can hold that power). A fixed range reports over-range if exceeded.
set_wavelength 🎛 control Set the correction wavelength (nm) the meter uses to convert sensor signal to power. Must be within the sensor's calibrated range (see get_device_info); sent as a whole number of nm.
zero_sensor 🎛 control Dark-zero the sensor (removes dark current / thermal offset). BEFORE calling, ask the user to block the beam and cover the sensor aperture completely, and wait for them to confirm: zeroing with light on the sensor corrupts all later readings. Takes a few seconds.

get_connection_info, get_command_log and reconnect are built into every LabMCP server.

A power meter only measures, so this server has no hazard tools. Your laser is the hazard: wear the right eyewear and never exceed the sensor's rated power or power density.

Safety limits

Limit Default Meaning
max_series_duration_s 600 s Longest power-logging series an agent may start, estimated as count × max(interval, reading time), where a reading takes averaging × 3 ms plus ~10 ms

Override at launch: --limit max_series_duration_s=3600. One call can log at most 3640 s (the tool's timeout minus a margin): longer estimates are refused, and a series that runs slower than estimated stops early and returns what it measured (completed: false). save_path must end in .csv and must not exist yet (an existing file is never overwritten); it is checked before logging starts.

Example prompts

  • "Set the meter to 1064 nm and tell me the power after the objective."
  • "Log the laser power every 2 s for 10 minutes and report the RMS and peak-to-peak stability. Save the data to ~/data/laser_warmup.csv."
  • "I've capped the sensor. Zero it, then I'll remove the cap and you read the power."
  • "Measure the transmission of this ND filter: read the power without it, then again after I insert it, and give me the optical density."
  • "Is the thermal sensor head still warming up? Read its temperature every minute until it's stable."

Notes

  • Wavelength matters. Photodiode responsivity varies by a factor of several across the sensor's range, so a wrong wavelength setting gives a wrong power. The server sends whole nanometres (the PM400/PM101/PM5020 references specify an integer), so 632.8 nm is set as 633 nm. That changes the responsivity by far less than the sensor's calibration uncertainty.
  • Units. MEAS:POW? answers in the meter's configured unit (W or dBm). The driver reads SENS:POW:UNIT? first and always returns watts, so a front-panel unit change can't mislead the agent.
  • Out of range. The meter answers INFINITY / 9.9E37 when the signal exceeds a manually selected range. The server reports this as an error that suggests auto-ranging.
  • Zeroing (SENS:CORR:COLL:ZERO) runs in the background. The server polls …:ZERO:STAT?, checks the error queue, and returns the zero value (…:ZERO:MAGN?). Zero values are lost when the meter reboots.
  • Energy (pyroelectric) sensors are recognised but not supported: single-pulse energy measurement needs a trigger/fetch workflow that this server doesn't implement yet.
  • Averaging is SENS:AVER:COUN. One sample takes ~3 ms on the PM100 series and 1 ms on the PM400/PM101/PM5020. The reply timeout scales with the averaging count.
  • PM101 and PM5020: Thorlabs' command references for these list exactly the commands used here. The PM5020 adds a channel suffix (SENS2, MEAS2, SYST:SENS2:IDN?), selected with --option channel=2. Neither has been tested yet, and PM5020 LAN/serial connections have not been tried. The PM100D's Windows TLPM DLL driver is not used; this server speaks SCPI only.
  • The simulator's error replies for out-of-range values (-222,"Data out of range") use the standard SCPI code. The Thorlabs documents only list -113,"Undefined header" explicitly.

Hardware verification

Model Firmware Interface Verified by Date
none yet: be the first

Facts

Kind
MCP server
Repo
K-Dense-AI/lab-instrument-mcps
Group
Uncategorized
Stars
19
License
Apache-2.0
Language
Python
Last push
2026-09-26
Forks
1
MCP Registry
io.github.K-Dense-AI/labmcp-thorlabs-pm
Homepage
www.k-dense.ai
Topics
ai-agents, instrument-control, lab-automation, laboratory, mcp, mcp-server, model-context-protocol, python, scientific-instruments, self-driving-labs

More on this shelf

  1. 1REAio.github.morluto/reaReverse engineer anything from your terminal or agent with one CLI and MCP server.125.3k
  2. 2com.browser-use/browser-usecom.browser-use/browser-useControl a real Chrome browser to complete any task: fill forms, extract data, book flights.117.7k
  3. 3World Monitorapp.worldmonitor/mcpLive markets, conflicts, country risk, chokepoints, energy, and China decision signals. 93 tools.88.3k
  4. 4Scrapling MCP Serverd4vinci/scrapling🕷️ An adaptive Web Scraping framework that handles everything from a single request to a full-scale crawl! Don't be shy, join here: https://discord.gg/EMgGbDceNQ and follow here for daily tips and tricks: https://x.com/Scraplingdev86.9k
  5. 5Everythingmodelcontextprotocol/serversThis MCP server attempts to exercise all the features of the MCP protocol. It is not intended to be a useful server, but rather a test server for builders of MCP clients. It implements prompts, tools, resources, sampling, and more to showcase MCP capabilities.85.8k
  6. 6Fetchmodelcontextprotocol/serversA Model Context Protocol server that provides web content fetching capabilities. This server enables LLMs to retrieve and process content from web pages, converting HTML to markdown for easier consumption.85.8k